Technical Data
Temperature – TC Type K (NiCr-Ni)
Measuring range |
-20 to +100 °C |
Accuracy |
±1 °C |
Resolution |
0.1 °C |
Reaction time |
< 50 s |
Temperature (ambient temperature)
Temperature – TC Type J (Fe-CuNi)
Measuring range |
0 to +400 °C |
Accuracy |
±1 °C (0 to +100 °C)
±1.5 % of mv (> 100 °C) |
Resolution |
0.1 °C |
Reaction time |
< 50 s |
Temperature (flue gas)
Flue gas O₂
Measuring range |
0 to 21 Vol.% |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
Reaction time t₉₀ |
30 s |
Flue gas Draught
Measuring range |
-20 to +20 hPa |
Accuracy |
±0.03 hPa (-3.00 to +3.00 hPa)
±1.5 % of mv (Remaining Range) |
Resolution |
0.01 hPa |
Flue gas degree of effectivity, Eta (calculated)
Measuring range |
0 to 99 % |
Resolution |
0.1 % |
Flue gas loss (calculated)
Measuring range |
0 to 99.9 % |
Resolution |
0.1 % |
Flue gas CO₂ calculation (calculated from O₂)
Measuring range |
0 to CO₂ max (Display range) |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
Reaction time t₉₀ |
< 40 s |
Pressure measurement
Measuring range |
-40 to +40 hPa |
Accuracy |
±0.5 hPa |
Resolution |
0.1 hPa |
Flue gas CO (without H₂-compensation)
Measuring range |
0 to 4000 ppm |
Accuracy |
±20 ppm (0 to 400 ppm)
±5 % of mv (401 to 2000 ppm)
±10 % of mv (2001 to 4000 ppm) |
Resolution |
1 ppm |
Reaction time t₉₀ |
60 s |
Ambient CO
Measuring range |
0 to 4000 ppm |
Accuracy |
±20 ppm (0 to 400 ppm)
±5 % of mv (401 to 2000 ppm)
±10 % of mv (2001 to 4000 ppm) |
Resolution |
1 ppm |
Reaction time |
60 s |
General technical data
Dimensions |
201 x 83 x 44 mm |
Operating temperature |
-5 to +45 °C |
Display type |
LCD |
Display function |
Backlit 2-line display |
Power supply |
Battery: 1500 mAh, mains unit 5V/1A |
Storage temperature |
-20 to +50 °C |
Weight |
(with probe) Approx. 660 g |